Ethylene oxide sterilization processes in the market differ in temperature, pressure, EO concentration and inertgas mixtures, i.e. CO2. Therefore the monitoring system should be validated with biological indicators before chemical indicators are used.For routine monitoring chemical indicators can be used with the advantage that they can be also checked immediately after the sterilization process has finished.The indicator system has been tested at 55°C under the following sterilization conditions:
During the sterilization process it is absolutely necessary to monitor the relative humidity, which should not be below 60%. The ideal relative humidity during sterilization is between 70% and 90%. A successful test confirms that the sterilizer meets the requirements of EN 1422 concerning air removal and ethylene oxide penetration.
Product Description
The Process Challenge Device (PCD) for ethylene oxide sterilization processes has been developed according to the standard EN 1422. It consists of a 4.55 m long stainless steel tube of 3 mm diameter connected with a capsule holding the indicator strip sealing one end.
Performance Characteristics
The PCD complies with a type test described in the European standard EN 1422, annex F, for air removal and ethylene oxide penetration.The combination of a PCD and a biological or chemical indicator is a class 2 indicator according to EN ISO 11140-
Operation Description
If all four bars of the chemical indicator turn from blue to green it is an indication of sufficient EOgas penetration inside the PCD. This result ensures air removal and EO-
Benefits
The use of this EO BMS allows the monitoring of sterility inside hollow instruments, tubes and porous goods not provided by recording pressure, temperature and EO concentration in the chamber and/or using exposed indicator strips.
The batch can be released without opening the pack to check the internal packing indicator.
All information relevant to release the load is supplied on completion of the process so that the person authorized can release the batch.
Cost effective. Only one indicator strip is required for each sterilization process instead of one in each pack.
Easy interpretation of the results due to precise colour change.
The graduated colour change of the indicator bars shows the level of the penetration into the PCD.
The indicator colour chemistry is a nonreversible chemical reaction. The indicator strip can be documented proof for several years without changing back to its original colour.
Environmentally friendly, no unnecessary waste.
gke self-
The test device can be used for an unlimitednumber of cycles. All important parts are made of stainless steel.
Continuous reproducibility of the results over the lifetime of the PCD.
All gke chemical indicators are protected from bleeding by a polymer binder and surface coating and can be disposed with normal garbage.
For validation biological indicators are available according to EN ISO 11138-
Assurance that only sterile released packs are used.